Installation
0112-0151 K
25
1.
Connect the quick connect fitting of the waste tubing to the
Waste Output fitting on the left side of the instrument.
2.
Connect the free end of the tubing to any vented waste
container.
For non-toxic waste the free end of the tubing can be placed
directly into a sink. The open end of the waste tubing must be
below the level of the Microplate Washer, and must remain free
of obstruction.
3.
If the bottle has a waste full sensor, connect the waste full sensor
to the Waste Sensor input fitting.
CAUTION!
An obstructed waste tubing can damage the
instrument.
WARNING! Biohazard. Waste tubing must be connected during
operation. Waste fluid is ejected out of the waste port. If no
waste tubing is connected, waste will spray and spill from the
waste output port.
CAUTION!
Waste containers must be vented. Connecting the
waste tubing to a pressurized or closed vessel will cause damage
to the waste pump.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...