AquaMax
®
2000/4000 Microplate Washer User Guide
150
0112-0151 K
Set
Response
:
“No data OK” response if the command succeeds.
Set
Example
:
/plate set 8 W0 B1 D5000 H10000
CR
returns OK
TAB
CRLF
>
Used
Action
:
Determines the total number of programs (protocols)
currently using the <index> plate.
Used
Response
:
Returns the number of programs.
Used
Example
:
/plate used 8 returns OK
TAB
1
CRLF
>
Auto
Action
:
Starts automatic well-bottom detection routine, if <type> is
given, or provides last determined well depth. The <type>
parameter describes the plate type, 0 for 96-well or 1 for
384-well.
Auto
Response
:
“No data OK” response if command succeeds. If <type> is
omitted, the last determined well depth is returned or 0 if
the procedure failed. The well depth value determined by
‘/plate auto’ is not stored in any plate index. Use ‘/plate set’
to store the well depth.
Auto
Example
:
/plate auto 0
CR
starts automatic detection and returns OK
TAB
CRLF
>
/plate auto
CR
returns OK
TAB
8512
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...