AquaMax Sterilant
0112-0151 K
117
Preparing the Reagent
The automated cleaning cycle on the AquaMax Microplate Washer liquid
handling system uses approximately 420 mL of Complete Sterilant.
Molecular Devices recommend that you prepare 500 mL of Complete
Sterilant to make sure that you have enough.
Preparing the Complete Sterilant
To prepare 500 mL of Complete Sterilant:
1.
Combine 125 mL of Sterilant Solution A Concentrate (4X) and
375 mL of deionized water to prepare 1X Solution A.
2.
Mix thoroughly.
3.
Within 1 hour of use, add 1 mL of Sterilant Solution B to the
500 mL of 1X Solution A.
4.
Mix thoroughly.
Using Reagent
For stubborn cleaning situations, Molecular Devices recommends using
the following procedure. You can modify the existing One-touch Cleaning
program or create a new program.
1.
Prime with Complete Sterilant (from the red inlet).
2.
Soak for 5 minutes, or up to 2 hours for endotoxin removal.
3.
Purge.
4.
Prime with deionized water (from the green inlet).
5.
Rinse with deionized water.
6.
Repeat rinse twice, for a total of three rinses.
7.
Purge.
Note:
You must use the Complete Sterilant within one hour of
preparation. Complete Sterilant cannot be stored for later use. However,
1X Solution A can be prepared up to 30 days in advance.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...