AquaMax
®
2000/4000 Microplate Washer User Guide
144
0112-0151 K
Repeat (REPE)
The protocol repeats steps from the specified first step to the repeat
command in order, the specified number of times. Firmware supports up
to 8 repeat loops per protocol.
Parameter tokens are:
Prime (PRIM)
Prime fills the internal fluid path with the specified fluid and removes
bubbles. This is similar to whole plate dispensing except that the fluid is
sent to waste and the wash vacuum valve is used to backfill the manifold.
Full prime is used when the system has been purged prior to prime.
Maintenance prime is used when changing inlets or other fluid is already
in the system.
Parameter tokens are:
A
Address (index) to repeat from [0-49]
N
Number of repetitions [1-99]
Example:
/STEP set 1 REPE A0 N2
CR
returns OK
TABCRLF
>
/STEP get 1
CR
returns OK
TAB
REPEAT A0 N2
CRLF
>
I
Inlet: [0-3], 0= green, 1= red, 2=blue, 3=yellow, -1= prompt
R
Rate: Prime type [0-1]: 0=Maintenance (~150 mL), 1 =full (~350 mL)
Example:
/STEP set 0 PRIM I2 R0
CR
returns OK
TABCRLF
>
/STEP get 0
CR
returns OK
TAB
PRIM I2 R0
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...