AquaMax
®
2000/4000 Microplate Washer User Guide
60
0112-0151 K
Prime Settings
The Prime step primes the internal fluidics and installed wash head with
the selected reagent. During the Prime step, the needles are immersed in
the priming liquid to wash the outsides of the needles. This is a
maintenance step and does not require a microplate to run.
Table 4-20:
Prime settings
Item
Description
1
Inlet–Liquid to be used for this step.
The available options correspond to the color-coded reagents
connected to the washer.
2
Prompt–Indicates that the system will pause before dispensing and
wait for the user to specify an inlet.
3
Mode–Level of priming fluid.
•
Maintenance: Uses 150 mL of the reagent to prepare the
instrument after long periods of system inactivity.
•
Full: Uses 400 mL of the reagent to prepare the instrument after
a wash head installation, reagent exchange, or system purge.
1
3
2
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...