Operation: The Touch Screen Interface
0112-0151 K
49
Dispense accuracy is set to specification during manufacturing. The
following factors can affect dispense accuracy:
•
Types of reagent.
•
Locations and size of reagent bottles.
The AquaMax Microplate Washer supports 4 L or 10 L bottles
only.
•
Level of reagent in bottles (too full or too low).
To adjust dispense factors, see
To adjust dispense accuracy for a program
Aspirate Settings
The Aspirate step removes fluid from a microplate using predefined
parameters.
Table 4-10:
Aspirate settings
Item
Description
1
Mode–Aspiration from Center, Edge, or Crosswise (from front to back
of wells).
Edge and Crosswise aspiration are supported for flat bottom wells
only.
2
Rate–Aspiration rate.
See
3
Descent speed–The speed at which the probes enter the wells.
4
Dwell time–Dwell time at the final probe position. The default value is
1 second. Range is 0-30 seconds.
1
4
3
2
5
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...