Rinse-flush (RFLU)
Rinse-flush step rinses all common fluidics (prime/soak trough, head,
valves) with fluid from the requested inlet. Upon completion the head is
left empty.
Parameter tokens are:
No Operation (NOP)
Skip the step.
Parameter tokens: none
Execute Step
Command: /
EXEC
[
PURGE
|
STAN
{
tokens
}]
I
Inlet: [0-3], 0= green, 1= red, 2=blue, 3=yellow, -1= prompt
Example:
/step set 2 RFLU i0
CR
returns OK
TAB
CRLF
>
/step get 2
CR
returns OK
TAB
RFLU i0
CRLF
>
Example:
/set 5 NOP
CR
, /get 5
CR
returns OK
TAB
NOP
CRLF
>
Action:
Execute a single step primitive without creating a protocol.
This command supports only purge and standby steps. The
PURGE and STAN options act identical to the corresponding
steps.