Robot Command Interface
0112-0151 K
145
Purge (PURG)
Purge removes all fluid from the internal fluidics path. The vacuum valves
are opened in the wash manifold and waste trough to remove all liquid.
Parameter tokens are: none
Standby (STAN)
Standby step fills the prime/soak trough with a rinse fluid and leaves the
dispense tips in the rinse fluid. At the specified refresh interval, the
trough is drained and refilled with the rinse fluid. When you stop the
operation normally (through a user interface selection), the trough is
drained and a small amount of fluid is rinsed (dispensed) through the
dispense needles.
Parameter tokens are:
Example:
/STEP set 1 PURG
CR
returns OK
TAB
CRLF
>
/STEP get 1
CR
returns OK
TAB
PURG
CRLF
>
I
Inlet: [0-3], 0= green, 1= red, 2=blue, 3=yellow, -1= prompt
T
Time: Refresh interval, 1-86400 [sec], (-1)= never
Example:
/ STEP set 2 STAN I2 T60
CR
returns OK
TAB
CRLF
>
/ STEP get 2
CR
returns OK
TAB
STAN I2 T7200
CRLF
>
Note:
This step primitive runs until “Stop” command is received. Upon
completion firmware drains the trough.
Note:
Firmware stores the standby interval as 16-bit word in minutes
(1-1440).
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...