Robot Command Interface
0112-0151 K
135
Rename
Action:
Renames Index protocol to <name>. One-touch protocols
0-2 cannot be renamed.
Rename
Response:
“No data OK” response if the command succeeds.
Get Action:
Reads a field, <m>, from the open protocol settings, where
<m> is 0 = start message, 1 = end message, 2 = protocol
parameters. A protocol must be opened (/protocol open)
before /protocol get can succeed.
Get Response:
Returns the start or end messages (<m>=0 or 1) or the
parameter list. The parameter list consists of single
character tokens followed by the token value, where the
tokens are:
P:
Plate ID, 0-31 or -1 for “no plate”.
F:
Dirty flag (128 = dirty, 0 = OK), indicating the
attached plate has changed since the protocol was
last stored.
M:
Stacker mode (0 = manual, 1 = automatic, use
stacker if attached).
R:
Restack mode (0 = no restacking, 1 = restack plates
after protocol completes with a stack of plates).
Get Example:
/prot get 0CRreturns OK
TAB
”Hello world”
CRLF
>
/prot get 2CRreturns OK
TAB
P2 F0 M0 R0
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...