![Molecular Devices AquaMax 2000 Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/molecular-devices/aquamax-2000/aquamax-2000_user-manual_1823779090.webp)
AquaMax
®
2000/4000 Microplate Washer User Guide
90
0112-0151 K
To recover the microplate and wash head manually
The Reset button on the bottom right corner of the Options tab performs
a software reset without power cycling the instrument. If the washer fails
during an operation, perform a reset to manually retrieve any microplate
still stuck in the system.
1.
To return the wash head and microplate to their unloaded
positions, touch
Reset
.
2.
Retrieve the microplate.
If the plate carriage cannot be unloaded by a software Reset,
continue with the following steps.
3.
Lift the wash head gently by the finger guard until it clears the
microplate.
4.
Disengage the lever mechanism and remove the wash head.
5.
Carefully slide the plate carriage out.
6.
If the plate carriage is stuck, touch
Load/Eject
in the
Options
tab.
7.
Remove the plate from the plate carriage.
CAUTION!
Do not touch or put pressure on the dispense or
aspiration probes, as they can be damaged easily.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...