Fast Wash (WHSQ)
Washing is sequential aspiration and dispensing. Aspiration at the
highest rate begins as with whole plate aspiration. After the aspiration
needles reach the specified height and the dwell period expires, the head
starts moving up and dispensing begins. The dispensing continues until
the specified volume per cycle is dispensed. Aspiration continues as the
needles move down to the specified height. Then the cycle repeats.
This mode does not support non-flat bottom plates or crosswise
aspiration.
Parameter tokens are:
I
Inlet: 0=green, 1=red, 2=blue, 3=yellow,
-1=prompt
R
Dispense rate [1 to 5]: 1=gentle, 5=fast
V
Dispense volume: 20-400 [
μ
L per well per
cycle]
H
Aspiration height from a plate bottom: [0
μ
m to
well depth]
U
Dwell time: Aspiration time0 to 25.5 [x 0.1 sec]
N
Number of cycles: [1 to 255]
Example:
/step set 3 SQWH 12 R0 V300 H1000 U0 N3
CR
returns
OK
TABCRLF
>
/step get 3 returns OK
TAB
SQWH 12 R0 V300 H1000 U0 N3
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...