Robot Command Interface
0112-0151 K
129
Program Control
Pause / Resume
Command: /
PAUSE
[no arg | 0 | 1]
Action:
Temporarily pauses (or resumes) execution of a running
program.
[no
arg]:
Reports current state.
[0]:
Resumes execution of a paused program.
[1]:
Requests a pause. The program is paused and the
status reports PAUSE only after the current step is
done. In PAUSE state, a program waits for the
“/pause 0" command from the host to proceed to
the next step.
Note:
Firmware switches to PAUSE state to display
start/stop program user messages and report
situations requiring user intervention. For example,
when a waste bottle is full. It expects to receive
“/pause 0” to resume or “/stop” to stop a program.
Response:
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...