Operation: The Touch Screen Interface
0112-0151 K
77
Table 4-26:
Options tab with administrative mode
Item
Description
1
Depressurize–Releases the system and bottle pressure. During
operation, the source bottles can be under mild pressure—up to 500
mmHg (10 psi or 70 kPa). Before changing the fluid in the source
bottles, depressurize to avoid releasing pressure when opening the
cap.
2
Purge–Purges the system fluidics. The wash head, prime trough, and
common fluid paths are evacuated. Use this function before changing
wash heads to remove excess fluids. In addition, after a system failure,
Purge can be used to drain the liquid left in the Prime/Soak trough.
3
Shutdown–Starts the user-defined automatic shutdown.
4
Admin–Starts and exits administrative mode and provides access to
secured functions. Administrative mode requires the administrative
passcode.
5
Load/Eject–Loads or ejects the plate carriage. Use this option to
retrieve a microplate after a system failure.
6
Reset–Performs a software reset without power cycling the
instrument.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...