![Molecular Devices AquaMax 2000 Скачать руководство пользователя страница 94](http://html1.mh-extra.com/html/molecular-devices/aquamax-2000/aquamax-2000_user-manual_1823779094.webp)
AquaMax
®
2000/4000 Microplate Washer User Guide
94
0112-0151 K
To prepare the wash head for storage
1.
On the
Main
tab, touch
Rinse
. Use AquaMax Sterilant, DI water,
or your usual rinsing agent.
2.
On the
Options
tab, touch
Purge
. The wash head can retain
approximately 3 cc of liquid after a purge.
3.
Release the locking lever on the wash head and lift the wash
head away from the instrument.
4.
Turn the wash head so that the inlets for the wash head are
pointing down over a folded paper towel. Avoid touching the
probes.
5.
Drain any remaining fluid from the wash head by gently tapping
the inlet end on the paper towel.
6.
Store the wash head in the provided container.
Note:
If you have followed the instructions in
, then skip
of this procedure
and start at
Tip:
To achieve the greatest level of dryness, you can do a
second rinse with ethanol through each of the four inlets on the
wash head to help dry the wash head and the system. Follow
, substituting ethanol as the rinse agent.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...