AquaMax
®
2000/4000 Microplate Washer User Guide
16
0112-0151 K
The plate library can store up to 32 plate types. For information about
adding a plate type, see
Managing the Plate Library on page 73
.
Reagent and Waste Bottle Ports
The source channels and waste connections are located on the left side
of the instrument. The AquaMax 2000 Microplate Washer has two
source channels, while the AquaMax 4000 Microplate Washer has four.
Each source channel can be connected to a reagent bottle, and is
composed of a fluid input, a pressure output, and an optional level
sensor connector. Waste can be drained into any waste bottle. A waste
sensor can be installed to detect a full waste bottle. Non-toxic waste can
be drained directly into a sink.
The AquaMax Microplate Washer supports 4 L or 10 L bottles only.
Table 1-5:
Reagent ports for AquaMax
®
4000 Microplate Washer.
Item
Description
1
Waste full sensor input
2
Waste fluid output
3
Source channels (green, red, blue, and yellow)*
4
Source bottle pressure outputs
5
Source bottle fluid inputs
6
Source bottle empty sensor inputs
1
2
6
5
4
3
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...