Maintenance
0112-0151 K
89
To remove the wash head
1.
To prepare the wash head for removal, run the one-touch
Rinse
program using sterile water as the rinse fluid. (See
2.
On the
Options
tab, touch
Purge
to remove water from the head.
3.
Disengage the lever mechanism and remove the wash head.
4.
Always store wash heads with the clear plastic guards installed,
with the probes facing down, and on a flat surface or inside a
protective container.
Note:
Repeat several times if a particularly harsh reagent was
used previously.
Note:
For long-term storage, use the clamshell container from
the original shipment.
Tip:
If the carriage does not eject automatically after an error condition
or power outage, retrieve the carriage manually by following
instructions in
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...