Operation: The Touch Screen Interface
0112-0151 K
81
Screen Calibration
The AquaMax Microplate Washer has a screen calibration feature that
enables the user to calibrate the position of the touch screen overlay on
the LCD screen.
To calibrate the screen
1.
On the
Options
Tab, touch
Calibrate Screen
.
Figure 4-9:
Screen calibration screen
2.
Touch the crosses as they appear on the screen.
The system exits Screen Calibration mode automatically when
the calibration is completed, or when the 10-second timeout is
reached.