Installation
0112-0151 K
21
Setting Up the Instrument
Set up the instrument on a level surface, away from direct sunlight, dust,
drafts, and vibrations. Complete the following procedures:
•
To turn on the instrument on page 21
•
To connect a reagent bottle on page 21
•
To connect a waste bottle on page 24
•
To install the wash head on page 26
To turn on the instrument
Molecular Devices recommends that you use a surge protector between
the power cord and the grounded power outlet.
1.
Connect the power cord to the back panel.
2.
Connect the power cord to a surge protector or grounded power
outlet.
The instrument automatically detects the voltage setting.
To connect a reagent bottle
Each reagent bottle comes with two quick-connect fittings and tubing in
two sizes. The larger tubing is for fluid. The smaller tubing is for bottle
pressure.
The AquaMax Microplate Washer supports 4 L or 10 L bottles only.
Note:
If a box has been damaged in transit, it is particularly important
that you retain it for inspection by the carrier in case there has also
been damage to the instrument.
WARNING! Each AquaMax Microplate Washer weighs approximately
41 lbs (18.5 kg) and should be lifted with care.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...