Operation: The Touch Screen Interface
0112-0151 K
69
8.
Weigh the plate after the dispense to measure the amount of
liquid dispensed.
9.
Divide the fluid mass by the fluid density and number of wells to
get the measured volume per well:
10.
Determine the fractional dispense accuracy using the equation
below:
11.
Round up this value to the nearest integer and type it in the
Adjust Dispense
dialog for the chosen fluid inlet. Use whole
numbers in the range of 50% to 200%.
12.
Repeat
for each fluid intake that needs
adjusting.
13.
Enter the results in the
Adjust Dispense
dialog of your working
program.
Measured volume per well
Fluid mass Fluid density
§
Number of wells (96 or 384)
---------------------------------------------------------------
=
Dispense accuracy
100
Measured Volume
Expected Volume
---------------------------------------
100
–
=
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...