Operation: The Touch Screen Interface
0112-0151 K
59
Standby Settings
Use Standby mode to keep wash fluid in the head without the need to
prime the head if there is a delay between uses.
The Standby step fills the prime/soak trough with a wash fluid and leaves
the dispense/aspirate probes in the wash fluid.
You can use the touch screen buttons to select the source bottle for the
wash fluid and to program the wash fluid to automatically change at
specific time intervals.
To discontinue the Standby step and continue using the AquaMax
Microplate Washer, touch Cancel. This drains the trough and dispenses a
small amount of the wash fluid through the dispense needles.
Table 4-19:
Standby settings
Item
Description
1
Inlet–Liquid to be used for this step.
The available options correspond to the color-coded reagents
connected to the washer.
2
Prompt–Indicates that the system will pause before dispensing and
wait for the user to specify an inlet.
3
Change solution every–Drains and refills the solution in the soaking
trough at the specified refresh interval.
1
3
2
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...