Robot Command Interface
0112-0151 K
139
Dispense (DISP)
Dispense step delivers a specified amount of fluid into the plate. The
dispense needles are moved to just above the top of the plate and the
requested horizontal position.The appropriate dispense valve is opened
for a calculated time to deliver the specified volume at the specified flow
rate.
For dispense rates available through the user interface, see
Plate Wash and Cell Wash Dispense Rates on page 48
Parameter tokens are:
M
mode
S
speed
A
address (index) of another step
N
number
I
inlet number, 0-3 (0=green, 1=red, 2=blue, 3=yellow), -1
(prompt)
R
Rate: Dispense rate (5 head specific pressure levels,
0=lowest, 4=highest, corresponding to the user interface
settings 1 to 5)
V
Volume: Dispense volume per well 20- 400 (96-well), 5-150
(384-well) [
μ
L]
Example:
/STEP set 5 DISP i0 r2 v350CR, response is OK
TABCRLF
>
/STEP get 5
CR
, response: OK
TAB
DISP I0 R2 V350
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...