AquaMax
®
2000/4000 Microplate Washer User Guide
126
0112-0151 K
Status
Command: /
STATUS
[no arg | y | z]
Action:
Provides information describing either the status of the
other instrument (no arguments) or the requested axis
status:
Response
IDLE
the instrument or AXTS is idle, no errors, ready to
run.
BUSY
nm
a protocol or a single command is being executed
(for example, axis is moving).
FAULT
nm
the instrument or AXTS cannot operate properly,
possible motion fault (axis stalled). It requires
RESET command to continue.
PAUSE
nm
an instrument or AXTS is paused waiting for a
user input - “Pause 0” to continue or “Stop” to
cancel or stop the action.
Response:
Where
<n> is a report flag: “1” = a new report message is available,
“0” = no news.
<m> is an index of the running protocol. A single-action
command in progress returns the step number as “-1”.
Example:
OK
TAB
IDLE
CRLF
>; OK
TAB
BUSY 0 3
CRLF
>; OK
TAB
BUSY 1 3
CRLF
>
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...