20.4.1.5
Reading out condition 4 (Host → device)
6
4
2
@
0
0
4
5
Terminator
Command No.
Address
00 ~ 31
Return (device → Host)
10
8
6
4
2
12
14
@
5
4
0
0
b
a
Terminator
Command No.
Address
00 ~ 31
Space
Status
a: Calibration mode [0] = DURING MEASUREMENT,
[1] = DURING ZERO CALIBRATION,
[2] = DURING SPAN CALIBRATION
b: Calibration error [0] = None
[1] = Occurrence of ERROR
20.4.1.6
Reading out ROM Version (Host → device)
6
4
2
@
0
0
4
6
Terminator
Command No.
Address
00 ~ 31
Return (device → Host)
10
8
6
4
2
12
14
@
6
4
0
0
V
✱
✱
✱
✱
Terminator
Command No.
Address
00 ~ 31
ROM version
*V1000 indicates ROM version 1.000.
20.4.1.7
Status change (Host → device)
Command No.
Operation
50
GROSS
51
NET
52
ZERO SET
53
ZERO CLEAR