[EFFECTIVE] is selected as default.
Note:
To execute the preset tare cancellation, please set the operation of the preset tare
weight cancellation to [EFFECTIVE] and set the preset tare weight amount in each brand.
9.11.2
Read-out operation of preset tare as zero (PRESET TARE=0 CHOICE)
The operation for the amount of preset tare is selectable from [THE LAST TARE IS USED]
or [OPERATION OF TARE CLR] in changing from the brand which set the preset tare to the
brand which does not set the preset tare under the condition that the operation of preset
tare weight cancellation is efective.
When [THE LAST TARE IS USED] is selected, the preset tare of the brand before the
change is applied. [TARE CLR] is selected as default.
Note:
-
When the tare weight cancellation is executed in the condition of the preset tare
being efective, the preset tare is cancelled and the tare weight cancellation is
executed.
The tare weight cancellation cannot be used together with the preset tare.
-
The tare is cleared if the zero set is executed in the condition of the gross weight
display.
9.12 Power-on-zero
When the power is turned on, this function [Power on ZERO] calibrats the display to zero
if the condition is stable, and the weight value is within ±10 % of the max. load (Max).
When the power is turned on, the function [Power ON ZERO] returns the compensation
processed by zero set and zero tracking to the weight value before calibrating the
compensation.
9.12.1
The operation of power–on-zero (ZERO AT POWER ON)
This function is selectable from [EFFECTIVE] or [INVALID].
[INVALID] is selected as default.
9.12.2
Clear of power–on-zero (ZERO CLR.POWER ON)
The operation of clear at power-on-zero is selectable from [EFFECTIVE] or [INVALID]
[INVALID] is selected as default.
9 Basic function setting display (BASIC FUNCTION)
Weight Controller CSD-918
Minebea Intec
EN-81