20.4.2 Data formats for (STREAM MODE), (SYNC. WITH PRINT), (SYNC. WITH ACCUM.), (SYNC.
WITH FINISH)
Note:
-
Load data is entered from the right.
-
[–] is entered for negative values and [+] for positive values.
-
Zero suppression is performed on the load data.
-
A decimal point is added at the speciied position if addition of decimal point is set
to [YES] and the decimal point is speciied on the calibration screen.
-
[OL] is displayed on the load data part when there is an overload.
-
The blank parts are all spaces.
20.4.2.1
When (TARGET OF STREAM/PRINT) is selected by (LOAD DISPLAY), (GROSS), or (NET)
Return (device → Host)
10
8
6
4
2
12
14
@
9
9
0
0
Terminator
Command No.
Load data
Load data sign
10
8
6
4
2
12
14
@
9
9
0
0
–
2
3
1
Terminator
10
8
6
4
2
12
14
@
9
9
0
0
+
2
3
1
.
0
Terminator
10
8
6
4
2
12
14
@
9
9
0
0
+
L
O
Terminator
10
8
6
4
2
12
14
@
9
9
0
0
–
L
O
Terminator
20 Appendix
Weight Controller CSD-918
Minebea Intec
EN-235
Содержание CSD-918 Series
Страница 239: ......