①
↓
-
Remove the connecting cable from the load cell terminal block.
-
Set the measurement range for the such measuring instruments as testers to V DC.
-
Measure the voltage across A-C after shorting A-F and C-G on the terminal block.
↓
↓
Is the voltage across
A and C stable at the
set excitation volta-
ge?
→ No → Contact Minebea Intec.
↓
Yes
↓
-
Reconnect the connecting cable to the load cell terminal block.
-
Check the load cell output voltage (refer to Chapter
).
↓
Is monitor display out
of ±3.1 mV/V range?
→ Yes → Shorten the terminal across B-
D in the connector.
↓
No
↓
↓
Recalibrate (see
Chapter
)
Is monitor dis-
play out of
±3.1 mV/V ran-
ge?
→ No →
↓
↓
Yes
↓
Is the status unchan-
ged?
→ Yes →
↓
No
↓
Start weighing
Provide
Minebea Intec
with details of
malfunction.
Consider the follo-
wing causes.
-
Input of load cell
is out of the ran-
ge of ±3.1 mV/V.
-
Load cell is bro-
ken due to over-
load, etc.
-
The signal cable
for the load cell is
disconnected.
Weight Controller CSD-918
19 Troubleshooting
EN-190
Minebea Intec