6.3.2
Change to weighing calibration screen
Please switch from weighing screen to weighing calibration screen by the following
operation.
6.3.3
Excitation voltage (BRIDGE POWER SUPPLY)
The excitation voltage (DC) for load cell can be set.
[12 V] is selected as default.
When CSD-918 is applied for the system where a zener barrier is used, select the
excitation voltage [BRIDGE POWER SUPPLY] properly according to the below table.
Excitation voltage (Bridge
power supply)
System coniguration
12 V DC 6 V DC
3 V DC
1 load cell + zener barrier
X
X
◯
2 load cells + cable junction box + zener barrier
X
X
◯
3 load cells + cable junction box + zener barrier
X
X
◯
4 load cells + cable junction box + zener barrier
X
X
◯
◯
: Applicable
X : Not applicable.
The above are the examples when 6-wire cable and 350
Ω
type of load cell are applied.
6.3.4
Set of scale interval (DIVISION)
Set the scale interval (min. scale division).
6.3.5
Set of max. load (Max) (WEIGHING CAPACITY)
Set the max. load (Max).
Note:
The display resolution that the performance becomes efective is 10 000 or less.
6.3.6
Zero calibration (ZERO ADJUSTMENT)
The input at zero point is registered.
There are two methods for zero calibration.
-
Calibration method by weight
1. Tap [MENU]
2. Tap [CALIBRATION] on the menu screen 1/2.
By tapping [BRIDGE POWER SUPPLY] in weighing calibration screen 2/2, select the
voltage from [12 V], [6 V] or [3 V].
Select [DIVISION] in weighing calibration screen 1/2, and select one from
[1],[2],[5],[10],[20] or [50].
Select [WEIGHING CAPACITY] in weighing calibration screen 1/2, and input any
arbitrarily numeric value in the range from [1] to [999 999].
Weight Controller CSD-918
6 Calibration
EN-42
Minebea Intec