4.4.4
Connecting the input
430 Ω 0.047μF
Approx.1
2 V 2.2 kΩ
A16, B16
A5~A6
B5~B8
COM 3
or
4.5 Connecting the external control I/O
This function facilitates external control using the external control I/O connector on the
rear panel. The external control input is executed by using a contact or open collector to
bridge each input and COM 1 after wiring the connector.
A1
A12
B1
B12
CONTROL I/O
The external control output is an open collector output (Open collector rating:
VCE = 35 V DC, IC = 40 mA DC at maximum).
The igure is an external control I/O connector.
The line from A1 to A12 is the input signal, and the line from B1 to B12 is the output signal.
4.5.1
Input connection
A1
Input 1
A2
Input 2
A3
Input 3
A4
Input 4
A5
Input 5
A6
Input 6
A7
Input 7
A8
Input 8
A9
Input 9
A10
Input 10
A11
Input 11
Approx. 12 V
2.2 kΩ
430 Ω
A1 to A11
or
A12
0.047 μF
COM. 1
Applicable connector:
FCN-361J024-AU; Cover FCN-360C024-B (Fujitsu)
A12
COM 1
4 Wiring
Weight Controller CSD-918
Minebea Intec
EN-31
Содержание CSD-918 Series
Страница 239: ......