17.1.5
Selection of target of analog output
The target of analog output is selected.
It is selectable from [LOAD DISPLAY], [GROSS] or [NET].
[LOAD DISPLAY] is selected as a default.
17.1.6
Scaling of analog output
The analog output is input with ten keys in the range from -999 999 to 999 999 to the
minimum value to maximum value in every [1] digit.
Set this as an optional value by switching from [ZERO OUTPUT] to[SPAN OUTPUT].
Analog output
Approx. 21.6 mA
Approx. 11 V
20 mA
10 V
Maximum display
4 mA
0 V
Minimum display
0 Display
Approx.2.4 mA
Approx. -1 V
Maximum display
Display
+110 % of the maximum display
-10 % of the maximum display
Note:
Equalize the scaling and output targets when a current output and a voltage output are
used synchronously.
(Fine calibration can be made to each output.)
The display when the minimum value of output is set to [ZERO OUTPUT].
The display when the maximum value of output is set to [SPAN OUTPUT].
Example:
[ZERO OUTPUT]: Set to 1 000.
[SPAN OUTPUT]: Set to 5 000.
Outputs maximum value at a display of 5 000.
Outputs minimum value at a display of 1 000.
3. Select [OPTION].
4. Select [ANALOG OUTPUT].