Weighing
Enforce within the [MAXTIME
OF COMPENSATION]
Time
Net
Waiting time for judge
Supplementary flow time
Waiting time until supplemen-
tary flow judgment
Stable
Finish
Over
OK
Under
Batch start
Final – Coarse Flow
Final – Middle Flow
Final – Fine Flow
Final
0
Batch start delay
Batch monitoring timer
Final – Under
Judge
Coarse flow
Middle flow
Fine flow
Set of control
Coarse Flow: Fixed open (weight)
Middle Flow: Fixed open (weight)
Fine Flow: Fixed open (weight)
11.6.4
Nozzle control sequence
This function is used when the matearial illed to the bottle through the inserted nozzle is
measured.
Before starting Flow, the output of nozzle DOWN turns on automatically, and the nozzle is
inserted into the container.
When the Dribble low has completed, the output of DOWN turns of and the inish of
measurement is judged with the nozzle pulled of.
During the nozzle DOWN (Waiting time of Flow start), when the Near zero is exceeded (the
case of the nozzle touched to the container is assumed), the sequence error [SQ.ERR 8] is
occurred and the control is stopped.
Weight Controller CSD-918
11 Weighing mode
EN-114
Minebea Intec