Note:
-
Be sure to bridge Sense+ and Sense- close to the load cell.
If bridged Sense+ and Sense- close to the device, the sense function will not work.
If the wire colors are diferent from those of Minebea Intec standard cables, refer
to the speciications of the load cell.
-
If the 6-wire cable length ≥100 m: the resistance of the cable causes the remote
sensing function to fail and accuracy may not be covered under warranty.
4.2.2
4-wire connection cable
Connect CSD-918 to a load cell using a 4-wire cable as follows.
Note:
The colors listed here apply for the Minebea Intec load cell and connection cables of
type "PR
"
Load cell
Additional bridges
A
+ EXC
Load cell power supply V+
red
F
+ SEN
Sense S+
white
G
– SEN
Sense S–
black
C
– EXC
Load cell power supply V–
blue
D
+ SIG
Load cell signal (meas. voltage) M+
green
4 Wiring
Weight Controller CSD-918
Minebea Intec
EN-25