6.7 CAL-switch
Lock calibration settings to protect them from changes. Use the CAL switch to lock the
calibration settings.
ON
OFF
3
2
3
1
To release the lock, switch the CAL switch OFF.
Note:
When the calibration is locked, the various setting listed below cannot be switched.
MENU
screen
BASIC FUNCTION
CALIBRATION
SERVO CALIBRATI-
ON
OPTION
A/D SAMPLING RA-
TE
ZERO ADJUSTMENT CONFIGURATION
ANALOG
OUTPUT
OPERATION OF
ZERO
SPAN ADJUSTMENT FULL OPEN TYPE1
RANGE OF ZERO
WEIGHING CAPACI-
TY
FULL OPEN TYPE2
TARGET OF ZERO
TRACKING
DIVISION
ZERO TRACKING
TIME
ZERO POINT mV/V
Each
setting
screen
ZERO TRACKING
BAND
SPAN POINT mV/V
1. Remove the (two) ixing screws (1) of the upper cover (2).
2. Slide the upper cover (2) toward the rear panel to remove it.
3. Switch the CAL switch (3) ON.
6 Calibration
Weight Controller CSD-918
Minebea Intec
EN-49