Set up an operation control panel if the instrument is used in the following locations:
-
Places exposed to water and/or spillages
-
Places contaminated with dust and/or ine particles
NOTICE
Risk of malfunctions
Use the instrument under the following conditions of temperature and humidity.
Environmental temperature: 0°C to 40°C.
Environmental humidity: Less than 85 % R. H. or less (Non condensing)
2.3.2
Installing the instrument
NOTICE
Warning of damage to property and/or the environment.
Install CSD-918 based on the following dimensions.
Ensure suicient space around the instrument.
The following are the dimensions of CSD-918 and the clearance required.
Front
192
96
CSD-918 Servo Weighing Controller
Rear
185
185
91
Side
131.5
(18.2)
20
Panel cut size
187
92
+1
0
+
1
0
all dimensions in mm
Side when servo control interface connector is installed.
The attached external I/O connector is the same size as the servo control interface
connector.
Weight Controller CSD-918