Check the RS-232C interface
↓
Is wiring correct?
→ No →
↓
Yes
↓
Connect wires surely with external
equipment (refer to Chap-
ter
).
Communication protocol is correc-
tly. set.
→ Yes →
↓
No
↓
Set it according to the host's speci-
ications (refer to Chapter
).
Command/data is set suitably.
→ Yes →
↓
No
↓
Use correct command/data (refer
to Chapter
).
Host speciies the low control.
→ Yes →
↓
No
↓
Switch the low control at host side
from [YES] to [NONE].
In Stream mode, intended value is
not sent.
→ Yes →
Select the output target in the
Stream mode.
↓
No
↓
Inform Minebea Intec about the
contents of failure and situation in
details.