(The input value of SPAN adjusted at the shipment, or calibrated at last time is
displayed.)
The SPAN registration is inished if the input value is reached below the [SPAN POINT
mV/V] in load calibration screen 1/2.
Note:
Error display of the span calibration
[SP-L]: Blinking for about 2 second when ([Load cell output voltage at SPAN] –
[Load cell output voltage at ZERO]) ≤ 0.0 mV/V (refer to Chapter
).
[SP-H]: Blinking for about 2 second when ([Load cell output voltage at SPAN] –
[Load cell output voltage at ZERO]) > 3.1 mV/V or when load cell output voltage exceeds
3.1 mV/V (refer to Chapter
Note:
The calibration accuracy obtained in Chapter
is less than 1/1000 or so.
When precise accuracy more than 1/1000 is necessary, please calibrate by "Calibration
method with weight".
6.4 Fine adjustment at Zero and SPAN (FINE ADJUSTMENT)
This function is ine calibration of zero and SPAN when there is the diferent margin in an
actual weight and the calibration weight.
6.4.1
Changeover to ine calibration screen
Please switch from weighing screen to weighing calibration screen by the following
procedure.
2. Please Input the diference of the load cell output value between equivalent to the
weighing capacity and equivalent to the zero point in mV/V.
3. Check that the content input with ten key is relected on the input check display.
4. Tap [OK].
1. Select [MENU].
2. Select [CALIBRATION] in menu screen 1/2.
3. Tap [NEXT] in load calibration screen 1/2.
4. Select [FINE ADJUSTMENT] in load calibration screen 2/2.
6 Calibration
Weight Controller CSD-918
Minebea Intec
EN-45