Analog output
Approx. 21.6 mA
Approx. 11 V
20 mA
10 V
Maximum display
4 mA
0 V
Minimum display
1000
Approx.2.4 mA
Approx. -1 V
5000
Display
5500
-10 % of the maximum display
Note:
-
The setting of [SPAN OUTPUT] must not exceed the max. load (Max), as described in
Chapter
-
Set [ZERO OUTPUT] to a lower value than [SPAN OUTPUT].
-
Equalize the scaling and output targets when a current output and a voltage output
are used synchronously.
Fine calibration can be made to each output.
17.1.7
Analog output ine calibration
After setting the scaling of analog output by [ZERO OUTPUT] and [SPAN OUTPUT],
proceed the ine calibration with the zero trimmer and the SPAN trimmer.
The ine calibration range is approx. ±1 % of full scale.
17.2 BCD output
17.2.1
Speciications (model CSD-918-P15)
The BCD output speciications are given below.
Speciications
I/O
Content
BCD data
Output 8 digit parallel output
POL. (polarity)
Output ON with the minus polarity output, and OFF with the
plus polarity output.
P.C. (print command)
Output ON in ixed time after conversion of BCD output is
completed.
Weight Controller CSD-918
17 Options
EN-142