A-13
Sensor type
Totally neonates
Data
Arms
518B
97 (51 male & 46 female)
200 pairs
2.38%
520N
122 (65 male & 57 female)
200 pairs
2.88%
The Pulse Oximeter with neonatal SpO2 sensors was also validated on adult subjects.
Refreshing rate
2 s
SpO
2
averaging time
7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Medium)
11 s (When the sensitivity is set to Low)
Masimo SpO
2
Module
SpO
2
Standards
Meet standards of ISO9919
Measurement range
1 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured without motion in neonate mode)
70 to 100%: ±3% (measured with motion)
1% to 69%: Not specified.
Refreshing rate
1 s
SpO
2
averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion SpO
2
accuracy ±2%
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...