A-23
Mainstream CO
2
Module
CO
2
Measurement range
0 to 150 mmHg
Accuracy
0 to 40 mmHg:
41 to 70 mmHg:
71 to 100 mmHg:
101 to 150 mmHg:
±2 mmHg
±5% of the reading
±8% of the reading
±10% of the reading
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours
Resolution
0 to 69 mmHg:
70 to 150 mmHg:
1 mmHg
0.25 mmHg
Rise time
<60 ms
awRR measurement range
0 to 150 rpm
awRR measurement accuracy
±1 rpm
Alarm limit
Range
Step
EtCO2 High
(low limit + 2) to 150 mmHg
EtCO2 Low
0 to (high limit - 2)mmHg
FiCO2 High
0 to 150 mmHg
1 mmHg
awRR High
(low limit + 2) to 100 rpm
awRR Low
6 to (high limit - 2) rpm
1 rpm
A.5.10 AG
Standards
Meet standard of ISO 21647 and YY0601
Technique Infrared
absorption
Warm-up time
Iso accuracy mode:
Full accuracy mode:
45 s
10 min
Sample flow rate
Adult, pediatric:
Neonate:
Accuracy:
120, 150, 200 ml/min
70, 90, 120 ml/min
±10 ml/min or ±10%, whichever is greater
Measurement range
CO
2
:
O
2,
N
2
O:
Des:
Sev:
Enf, Iso, Hal:
awRR:
0 to 10%
0 to 100%
0 to 18%
0 to 8%
0 to 5%
2 to 100 rpm
Resolution
CO
2
:
awRR:
1 mmHg
1 rpm
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...