11.2 Safety
WARNING
z
Use only SpO
2
sensors specified in this manual. Follow the SpO
2
sensor’s
instructions for use and adhere to all warnings and cautions.
z
When a trend toward patient deoxygenation is indicated, blood samples should be
analyzed by a laboratory co-oximeter to completely understand the patient’s
condition.
z
Do not use SpO
2
sensors during magnetic resonance imaging (MRI). Induced
current could potentially cause burns.
z
Inspect the sensor site every two hours and move the sensor if the skin quality
changes. For neonates, or patients with poor peripheral blood circulation or
sensitive skin, inspect the sensor site more frequently. Prolonged continuous
monitoring may increase the risk of undesirable changes in skin characteristics,
such as irritation, reddening, blistering or necrosis.
11.3 Identifying SpO
2
Modules
To identify which SpO
2
module is incorporated into your patient monitor, see the company
logo located at the lower left corner on your monitor’s front. If it is:
Mindray SpO
2
module, there is no logo.
Masimo SpO
2
module, there is the Masimo SET logo.
Nellcor SpO
2
module, there is the Nellcor logo.
The connectors for these three SpO
2
sensors are mutually exclusive.
11.4 Applying the Sensor
1. Select an appropriate sensor according to the module type, patient category and weight.
2. Remove colored nail polish from the application site.
3. Apply the sensor to the patient.
4. Select an appropriate adapter cable according to the connector type and plug this cable
into the patient monitor.
5. Connect the sensor cable to the adapter cable.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...