15-6
15.6 Changing C.O. Settings
15.6.1 Setting the Temperature Unit
In the [
C.O. Setup
] menu, select [
Temp Unit
] to toggle between [
ºC
] and [
ºF
].
15.6.2 Setting the Interval between Measurements
To avoid inaccurate measurements, a certain period of time should be allowed for the blood
temperature to become stable before starting a new measurement. To set the interval
between two measurements, in the [
C.O. Setup
] menu, select [
Interval (s)
] and enter an
appropriate value. Every time when a measurement is completed, the monitor will count
down the remaining time and give corresponding prompt messages. A new measurement
will not start until the time is elapsed.