5-6
To set the default configuration at startup,
1. Enter the configuration mode.
2. In the load configuration area, select a factory default configuration of the current
department or a user configuration as the default configuration.
3. Select
[
Adu
] or [
Ped
] or [
Neo
] from the [
Patient Cat.
] as the default patient category.
4. You can change the department by selecting [
Change Department >>
]. But this will
delete all the saved user configurations.
5. Set
the
[
Load Selected Config at Startup
] option:
Selected: When the monitor restarts after quitting for over 120 seconds, it will
load the configuration you have selected in the load configuration area.
Deselected: The monitor will load the latest configuration at startup no matter
how long it has been quitted.
NOTE
z
To find out which configuration has been loaded when the monitor starts, check
the message (lasting for about 10 seconds) displayed at the bottom of the screen
after the main screen comes out.
5.8 Loading the Latest Configuration Automatically
During operation, you may make changes to some settings. These changes may not be saved
as user configuration. To prevent the changes from loss in case of a sudden power failure,
the monitor stores the configuration in real time. The stored configuration is the latest
configuration.
If [
Load Selected Config. at Startup
] is selected, the monitor will load the latest
configuration if restarts within 60 seconds after the power failure. And it will load the
default configuration rather than the latest configuration if restarts 120 seconds later after
the power failure. The monitor may load either the latest configuration or the default
configuration if restarts from 60 to 120 seconds after the power failure.
If [
Load Selected Config. at Startup
] is not selected, the monitor will load the latest
configuration at startup.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...