26-6
6. Enter the vented gas concentration. If you use only one gas for calibration, set other
gases’ concentration to 0.
7. Select
[
Start
] to start calibration.
8. If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
AG Cal. Failed
] is displayed. Perform
another calibration.
9. Select
[
Exit
] to exit the current menu.
26.8 Setting up IP Address
1. In the main menu, select [
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password and then select [
Network Setup >>
] from the popup menu.
2. If your monitor is equipped with a wireless AP, you can set [
Network Type
] to [
WLAN
]
in the network setup menu. Otherwise, the default setting is [
LAN
].
3. Set the IP Address and subnet mask through [
IP Address
] and [
Subnet Mask
].
If the patient monitor is connected to a CMS, its IP address should be set up. The user should
not change the patient monitor’s IP address randomly. If you want to know details about IP
address setup, contact the technical personnel in charge of the CMS.
26.9 Entering/Exiting Demo Mode
To enter the Demo mode:
1. In the main menu, select [
Maintenance >>
].
2. Select
[
Demo >>
]. Enter the required password and then select [
Ok
].
To exit the Demo mode:
1. In the main menu, select [
Maintenance >>
].
2. Select
[
Exit Demo
] and then select [
Ok
].
3. The patient monitor exits the Demo mode.
WARNING
z
The Demo mode is for demonstration purpose only. To avoid that the simulated
data are mistaken for the monitored patient’s data, you must not change into
Demo mode during monitoring. Otherwise, improper patient monitoring and
delayed treatment could result.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...