A-10
HR averaging method
In compliance with the requirements in Clause 4.1.2.1 d) of
ANSI/AAMI EC13, the following method is used:
If the last 3 consecutive RR intervals are greater than 1200 ms, the 4
most recent RR intervals are averaged to compute the HR. Otherwise,
heart rate is computed by subtracting the maximum and minimum ones
from the most recent 12 RR intervals and then averaging them.
The HR value displayed on the monitor screen is updated every second.
Response to irregular
rhythm
In compliance with the requirements in Clause 4.1.2.1 e) of
ANSI/AAMI EC13, the heart rate after 20 seconds of stabilization is
displayed as follows:
Ventricular bigeminy (3a): -80±1 bpm
Slow alternating ventricular bigeminy (3b): -60±1 bpm
Rapid alternating ventricular bigeminy (3c): -120±1 bpm
Bidirectional systoles (3d): -90±2 bpm
Response time to heart
rate change
Meets the requirements of ANSI/AAMI EC13: Section 4.1.2.1 f).
From 80 to 120 bpm: less than 12 s
From 80 to 40 bpm: less than 12 s
Meets the requirements of ANSI/AAMI EC13: section 4.1.2.1 g).
Time to alarm for
tachycardia
Waveform
4ah - range:
4a - range:
4ad - range:
Waveform 4bh - range:
4b - range:
4bd - range:
11 s
11 s
11 s
11 s
11 s
11 s
Tall T-wave rejection
capability
Meets the requirements of part 4.1.2.1 c) of ANSI/AAMI EC 13 The
heart rate meter rejects all 100 ms QRS complexes with less than 1.2
mV of amplitude, and T waves with T-wave interval of 180 ms and
those with Q-T interval of 350 ms.
Arrhythmia Analysis
Classifications
Asystole, VFib/VTac, Couplet, Bigeminy, Trigeminy, R on T, VT>2,
PVC, Tachy, Brady, Missed Beats, PNP, PNC
ST Segment Analysis
Measurement range
-2.0 to 2.0 mV
Accuracy
-0.8 to 0.8 mV:
Beyond this range:
±0.02 mV or ±10%, whichever is greater.
Not specified.
Refreshing rate
10 s
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...