7-7
7.5.2 Displaying Alarm Limits
To have a better view of measurement numerics, you can choose not to display alarm limits
in the parameter window by switching off displaying alarm limits in configuration mode.
For details, refer to the chapter
Managing Configuration
.
7.5.3 Setting Alarm Delay Time
You can set the alarm delay time for over-limit alarms of continuously measured parameters.
If the alarm triggered condition disappears within the delay time, the patient monitor will
not issue the alarm. You can select [
Alarm Setup>>
] from [
Main Menu
] and then set
[
Alarm Delay
].
In the configuration mode, you can follow these steps to set [
Alarm Delay
] and [
ST Alarm
Delay
] respectively.
1. Press
the hardkey on the monitor’s front to enter the main menu.
2. Select
[
Manage Configuration >>
]
→
[
Configuration >>
]. Enter the required
password and then select [
Ok
].
3.
Select
[
View Config
.]
→
[
Alarm
] and set [
Alarm Delay
] and [
ST Alarm Delay
].
NOTE
z
The alarm delay time can not be set in case of the following physiological alarms:
Resp Apnea (The apnea time needs to be set independently), ST Alarm (The delay
time needs to be set independently), Arrh. Alarm, PVCs Too High, ECG Weak
Signal,
Resp Artifact,
SpO
2
Desat,
No pulse,
Nellcor SpO
2
over-limit alarm, FiO
2
Too Low, over-limit alarms of discontinuously measured parameters and MAC >3.
7.5.4 Using Alarms
Switching Individual Measurement Alarms On/Off
1. Select the parameter window for your desired measurement to enter its setup menu.
2. Select
[
Alarm
] and toggle between [
On
] and [
Off
].
[
On
]: The patient monitor gives alarm indications in accordance with the preset
alarm level and stores the related waveforms and parameters.
[
Off
]: The alarm off symbol “
” is displayed in the measurement parameter
window.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...