8-21
8.8 12-Lead ECG Monitoring
8.8.1 Entering the 12-lead ECG Monitoring Screen
1. Refer to the section
8.3.3 ECG Lead Placement
for placing the electrodes.
2. Select the ECG parameter window and then set [
Lead Set
] to [
12-Lead
], and [
ECG
Display
] to [
12-Lead
] in the popup menu.
There are totally 12 ECG waves and 1 rhythm wave displayed on the screen. The rhythm
lead is the HR-derived lead before entering the 12-lead ECG monitoring screen. The ST
numerics are displayed in three groups:
ST Ant (anterior): V1, V2, V3, V4
ST Inf (inferior): II, III, aVF, (aVR)
ST Lat (lateral):
I, aVL, V5, V6
Although aVR is displayed in the ST Inf group, it is not an inferior lead.
Additionally, the 12-lead ECG monitoring has the following features:
The [
Filter
] mode is fixed to [
Diagnostic
] and cannot be changed.
The
hardkey on the monitor’s front is disabled.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...