26-2
26.2 Service Tasks
The following tasks are for our qualified service professionals only. Contact a qualified
service provider if your patient monitor needs the following services. Clean and disinfect
equipment to decontaminate it before testing or maintaining it.
Maintenance schedule
Frequency
Safety checks according to
IEC60601-1
At least once every two years, after any repairs where the
power supply is replaced, or if the patient monitor has been
dropped.
Performance assurance for all
measurements not listed below
At least once every two years, or if you doubt the measured
values.
Defibrillator synchronization
At least once every two years, or as needed.
NIBP leakage test
At least once every two years, or by your hospital policy.
NIBP accuracy test
At least once every two years, or by your hospital policy.
NIBP calibration
At least once every two years, or by your hospital policy.
CO
2
calibration and performance
test
At least once a year, or if you doubt the measured values.
AG calibration
At least once a year, or if you doubt the measured values.
AG preventative maintenance
(pump check, internal fan check,
ambient pressure check and so forth
as described in the service manual)
At least once a year, or if you doubt the measured values.
26.3 Checking Monitor and Module Information
In the main menu, select [
Maintenance >>
]
→
[
System Status >>
]. You can view the
information about system start time, selftest, etc. from the popup menu. You can print out the
information for the convenience of troubleshooting. The information will not be saved during
shut down.
You can also view the information about the monitor configuration and system software
version by selecting [
Maintenance >>
]
→
[
Software Version >>
] in the main menu.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...