16-3
CAUTION
z
The watertrap collects water drops condensed in the sampling line and therefore
prevents them from entering the module. If the collected water reaches a certain
amount, you should drain it to avoid blocking the airway.
z
The watertrap has a filter preventing bacterium, water and secretions from
entering the module. After a long-term use, dust or other substances may
compromise the performance of the filter or even block the airway. In this case,
replace the watertrap. It is recommended to replace the watertrap once every two
months, or when the watertrap is found leaky, damaged or contaminated.
16.2.2 Using a Microstream CO
2
module
1. Connect the sampling line to the module and then connect the CO
2
components as
shown below.
2. By default, the microstream CO
2
module is in measure mode. The message [
CO
2
Sensor Warmup
] appears on the screen when the CO
2
module is plugged.
3. After warm-up, you can perform CO
2
measurements.
Sampling line
Connector for sampling line
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...