Maintenance
WARNING
z
Failure on the part of the responsible individual hospital or institution employing
the use of this equipment to implement a satisfactory maintenance schedule may
cause undue equipment failure and possible health hazards.
z
The safety checks or maintenance involving any disassembly of the equipment
should be performed by professional servicing personnel. Otherwise, undue
equipment failure and possible health hazards could result.
z
If you discover a problem with any of the equipment, contact your service
personnel or us.
26.1 Safety Checks
Before every use, after your patient monitor has been used for 6 to 12 months, or whenever
your patient monitor is repaired or upgraded, a thorough inspection should be performed by
qualified service personnel to ensure the reliability.
Follow these guidelines when inspecting the equipment:
Make sure that the environment and power supply meet the requirements.
Inspect the equipment and its accessories for mechanical damage.
Inspect all power cords for damage, and make sure that their insulation is in good
condition.
Make sure that only specified accessories are applied.
Inspect if the alarm system functions correctly.
Make sure that the recorder functions correctly and the recorder paper meets the
requirements.
Make sure that the batteries meet the performance requirements.
Make sure that the patient monitor is in good working condition.
Make sure that the grounding resistance and leakage current meet the requirement.
In case of any damage or abnormity, do not use the patient monitor. Contact the hospital’s
biomedical engineers or your service personnel immediately.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...