3-11
3.9.1 Setting up a Monitor
In situations where you install a patient monitor or change the patient monitor’s application
site, you need to setup the patient monitor as follows:
1. In the main menu, select [
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password.
2. Select
[
Device Setup>>
]. In the pop-up menu, select, in turn, [
Monitor Name
],
[
Institution Name
], [
Department
] and [
Bed No.
], and then change their settings.
3.9.2 Changing Language
1. In the main menu, select [
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password and then select [
Device Setup >>
].
2. In
the
[
Device Setup
] menu, select [
Language
] and then select the desired language.
3. Restart the patient monitor.
3.9.3 Setting DIAP Baud Rate
1. In the main menu, select [
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the
required password and then select [
Device Setup >>
].
2.
Select
[
DIAP Baud Rate
] and then select the desired baud rate.
WARNING
z
When connecting compatible devices to the patient monitor via RS232 serial port,
make sure the baud rate of the compatible devices is consistent with that of the
patient monitor. Otherwise, the compatible devices can’t communicate with the
patient monitor.
3.9.4 Adjusting the Screen Brightness
1. In the main menu, select [
Others >>
].
2. Select
[
Brightness
].Select the appropriate setting for the screen brightness. 10 is the
brightest, and 1 is the least bright. If the patient monitor operates on battery power, you
can set a less bright screen to prolong the operating time of the battery.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...