D-6
Measurement
Alarm message
L
I
Cause and solution
NIBP Illegally Reset L
C
An illegal reset occurred during NIBP
measurement. Check if the airway is occluded.
IBP YYSensor
Off
(YY represents an
IBP label.)
L
A
C.O.
TB Sensor Off
L
A
Check the sensor connection and reconnect the
sensor.
CO2 Sensor High
Temp
L
C
Check, stop using or replace the sensor.
CO2 Sensor Low
Temp
L
C
Check, stop using or replace the sensor.
CO2 Airway High
Press.
L
C
CO2 Airway Low
Press.
L
C
An error occurred in the airway pressure. Check
the patient connection and patient circuit, and then
restart the monitor.
CO2 High
Barometric
L
C
CO2 Low
Barometric
L
C
Check the CO2 connections, make sure that the
monitor application site meets the requirements,
and check for special sources that affect the
ambient pressure. Restart the monitor.
CO2 FilterLine
Occluded
L
C
The airway or watertrap was occluded. Check the
airway and remove the occlusion.
CO2 No Watertrap
L
B
Check the watertrap connections.
CO2 Check Adapter
L
C
There is a problem with the airway adapter.
Check, clean or replace the adapter.
CO2 FilterLine Err
L
C
Check if there is a leak in the CO
2
sample line or
the CO
2
sample line has been occluded.
CO2 Zero Failed
L
C
Check the CO
2
connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO2 Check Cal.
L
C
Perform a calibration.
CO2 Check Airway
L
C
An error occurred to the airway.
CO2 No Filterline
L
A
Make sure that the filterline is connected.
CO2 Main Board
Err
H
C
CO2 Replace
Scrubber&Pump
L
C
CO2 15V Overrange H
C
CO2 Hardware Err
H
C
There is a problem with the CO
2
module. Restart
the monitor.
CO
2
CO2 User Cal Fail
L
C
Re-calibrate the AG module.
Содержание iPM-9800
Страница 1: ...iPM 9800 Patient Monitor Operator s Manual ...
Страница 2: ......
Страница 44: ...3 14 FOR YOUR NOTES ...
Страница 50: ...4 6 FOR YOUR NOTES ...
Страница 58: ...5 8 FOR YOUR NOTES ...
Страница 102: ...8 24 FOR YOUR NOTES ...
Страница 108: ...9 6 FOR YOUR NOTES ...
Страница 118: ...11 8 FOR YOUR NOTES ...
Страница 128: ...12 10 FOR YOUR NOTES ...
Страница 137: ...14 7 NOTE z After entering the PAWP measurement window the monitor will turn off the PA alarm automatically ...
Страница 140: ...14 10 FOR YOUR NOTES ...
Страница 166: ...17 10 FOR YOUR NOTES ...
Страница 178: ...19 10 FOR YOUR NOTES ...
Страница 188: ...20 10 FOR YOUR NOTES ...
Страница 200: ...22 6 FOR YOUR NOTES ...
Страница 206: ...23 6 FOR YOUR NOTES ...
Страница 212: ...24 6 FOR YOUR NOTES ...
Страница 260: ...A 28 FOR YOUR NOTES ...
Страница 266: ...B 6 FOR YOUR NOTES ...
Страница 299: ......
Страница 300: ...PN 9211 20 87439 7 0 ...